Fault Treatment Method of Spring Testing Machine


Release Time:

Dec 21,2023

1999 this happens when the electronic testing machine is turned on, the CPU working frequency (11.0592MHz) may be abnormal, and the crystal oscillator will be replaced on the same day. If the input terminal voltage of the two sensors is too high (usually a few millivolts), check whether the sensors are normal. The general data obtained by electrically blocking the measurement sensor with a multimeter are: the input terminal is about 405Ω; the output terminal is about 340Ω; the resistance value of the two leads at either end and the other end is 280 Ω. Note: the sensor input, the output of the data sometimes with different manufacturers will have a certain change, but the change is a certain proportion. If the comparison voltage of the two operational amplifiers is not normal, measure whether the resistance of pin 2 and pin 3 of the amplifier has changed. If the resistance value occurs, it should be replaced. If the above conditions are normal, the operational amplifier of the electronic testing machine is damaged and should be replaced. If the 220V voltage of the power supply is unstable, a regulated power supply shall be configured.

Fault Treatment Method of Spring Testing Machine

1999 this happens when the electronic testing machine is turned on, the CPU working frequency (11.0592MHz) may be abnormal, and the crystal oscillator will be replaced on the same day. If the input terminal voltage of the two sensors is too high (usually a few millivolts), check whether the sensors are normal. The general data obtained by electrically blocking the measurement sensor with a multimeter are: the input terminal is about 405Ω; the output terminal is about 340Ω; the resistance value of the two leads at either end and the other end is 280 Ω. Note: the sensor input, the output of the data sometimes with different manufacturers will have a certain change, but the change is a certain proportion. If the comparison voltage of the two operational amplifiers is not normal, measure whether the resistance of pin 2 and pin 3 of the amplifier has changed. If the resistance value occurs, it should be replaced. If the above conditions are normal, the operational amplifier of the electronic testing machine is damaged and should be replaced. If the 220V voltage of the power supply is unstable, a regulated power supply shall be configured.

If the amplifier output voltage is not stable, replace the amplifier. Check whether the CPU frequency (11.0592MHz) and the comparison voltage are stable or drift? If there is a virtual welding problem, it should be re-welded; if the performance of the components is poor, it should be replaced to check whether the circuit board is damp or oily, and necessary cleaning and drying work should be done. If the power transformer has no output, replace the transformer or rewind the transformer. If the voltage stabilizing circuit of the electronic testing machine is abnormal, check the corresponding bridge diode, filter capacitor and corresponding connecting line. Replace the damaged ones, and reweld the damaged ones. Check that the digital circuit voltage is +5V.

Key words:

Consulting

Note: Please leave your contact information and our professionals will contact you as soon as possible!